Thoracic Pedicle Screw Placement Using the Outside-In Technique: An Analysis Of Screw Placement Using Anatomic Landmarks Without Image Guidance

Information provided by

Exhibit from the SRS 2002 Annual Meeting
PURPOSE: To evaluate the safety of the “outside-in” technique utilizing only standard anatomic landmarks without image guidance.

MATERIALS & METHODS: 80 pedicle screws were placed from T4-T11 in 5 fresh cadaveric specimens. No imaging was utilized. All screws were placed using anatomic landmarks. Pedicles were dissected to determine breakout, and location of anatomic structures. Breakout was described as Grade 1 <1mm ,Grade II <2m, Grade III 2-4mm of screw penetration

RESULTS: 3 screws (3.75%) perforated the medial cortex: 2 Grade I, 1 Grade II. 4 screws (5%) violated the pedicle directly lateral: 2 Grade I, 1 Grade II, 1 Grade III. The rib laterally covered all screws. There were no pedicle violations superior or inferior.

CONCLUSIONS: The “outside-in” technique for thoracic pedicle screw placement is safe and utilizes the same anatomic landmarks at all levels. Our results are comparable to other published studies where image guidance was utilized.

Updated on: 12/10/09