The Effects of Washers and Staples on Anterior Spinal Instrumentation Stability: Can Plowing be Prevented? A Cadaveric Model
Information provided by

Poster from the SRS 2002 Annual Meeting
The purpose was to determine if washers and staples used in anterior
spinal instrumentation provide a mechanical advantage
in eliminating plowing of the screws. Before testing they were
separated into 15 single vertebra, potted, and instrumented
with a screw, screw-washer, or screw staple equally. A rod was
attached and axial compression was applied. Results were
compared to ASTM UHMW models. Sample t-test assuming unequal variances
yielded that washer vs. screw was not
significantly different (p > 0.176), staple vs. screw was not
significantly different (p >0.145), and staple vs. washer was
not significantly different (p > 0.678). Results were identical
to tests using the ASTM model. Washers and staples did not provide
a mechanical advantage or make a significant contribution to the
stability of anterior scoliosis constructs. Staples actually
slightly decreased the stiffness of the construct compared to
the screw only construct. Plowing of the screws was not affected
by washers or staples.
Posted on: 02/10/04 | Updated on: 12/10/09
Related Articles
- Genetically Modified Human Derived Bone Marrow Cells for Postero-Lateral Lumbar Spine Fusion in Athymic Rats
- Severe Infantile Scoliosis Treated with Repetitive Distractions Followed by Definitive Arthrodesis
- Biomechanical, Radiographic, and Histological Healing Characteristics of Anterior Spinal Fusion in a Sheep Model
- Treatment of Degenerative Disc Disease and Degenerative Spondylolisthesis of the Lumbar Spine - Figures 4 a-e
- Treatment of Degenerative Disc Disease and Degenerative Spondylolisthesis of the Lumbar Spine - Figure 1 a-b
- Treatment of Degenerative Disc Disease and Degenerative Spondylolisthesis of the Lumbar Spine - Combinations